Receptacle Configuration
5-20R
Standards
ANSI/IEEE: Category A, NFPA: NFPA 99 2012 ed, UL 1363A and CSA C22.2: No. 21, UL 1449 3rd Ed and CSA C22.2: No. 8, UL 60601-1 and CSA C22.2: No.60601-1, UL 60950-1 and CSA C22.2: No. 60950-1
Noise Attenuation
-6 to -29dB
Mounting Style
Mounting Hole
Special Features
Easily Accommodates Surface Mounting and IV Pole Mounting With the Use of the Mounting Bracket, Heavy Duty Steel Construction Withstands the Rigors of Demanding Environments, Locking Safety Covers are Splash-Resistant and Protect From Incidental Contact With Unused Outlets
Application
Hospital and Medical Facility
Medical Facility Use
Outside Patient Care Areas (UL 1449); Patient Care - Only When Permanently Mounted (UL 1363A); Patient Care (UL 60601-1)
For Use With
Isolated Power Systems
Surge Protection Indicator
Yes
UL 1363A u0026 CSA C22.2
No. 21
Enclosure
Powder Coated Steel
Outlet Covers
Polypropylene
Noise Filtering
-6dB to -29dB (100kHz to 100MHz)
Maximum Surge Current, Per Mode (L-N)
27kA
UL 60950-1 u0026 CSA C22.2
No. 60950-1
HG Outlet Configuration
NEMA 5-20R
Maximum Leakage Current
u003c100µA
Voltage Protection Rating (VPR) L-N, L-G, N-G
600V, 600V, 500V
Operating Temperature
0°C to 40°C
Country of Origin (subject to change)
Mexico